Abstract

An implantable joint angle transducer (IJAT) was developed to provide command-control and feedback-control information for chronic use with functional neuromuscular stimulation (FNS) neuroprostheses. The IJAT uses Hall effect sensors to transduce joint angle. A titanium encapsulated array of Hall effect sensors and support circuitry is surgically implanted in one bone, and a similarly encapsulated permanent magnet in an opposing bone, across a joint. The IJAT provides consistent, reliable, high quality signals that reflect joint movement from midsized two-degree-of-freedom joints. IJAT's were implanted using a chronic in vivo dog model to demonstrate the feasibility of implantation and periodic measurement techniques, and to validate modeling techniques used for prediction of function and calibration. The flexion resolution ranged from 0.4 to 3.0 degrees over a range of 115 degrees. The maximum deviation from a linear response was 9 degrees. The resolution and linearity depend on several transducer and joint geometry parameters, and can be predicted prior to implantation and calibrated after implantation. The results of this study 1) defined the most appropriate hermetic capsule designs for the IJAT sensor and magnet, 2) defined the best orientation of the magnetic field to optimize device function, 3) provided a computer model of the IJAT to aid in placement, calibration, and evaluation of the device, 4) verified the surgical techniques used to implant the device, and 5) verified the long-term functionality and the biocompatibility of the device.