Abstract

We prepared an implantable device of small volume (SID) that is enabled with on-demand, pulsatile drug release. The device was designed to be actuated via a magnetic field; hence, there was no need for a battery. The device was actuated when the magnet was applied from the outside and infused the drug solution outward via the outlet ports in the device. When there was no external magnetic field, no drug was released. In this work, we varied the amount of delivered drug by varying the number of outlet ports. Thus, as the number of outlet ports increased from one to three, the average amount of drug release per actuation increased from 60.7 ± 1.79 μg to 122.6 ± 1.27 μg. In addition, when the SID with three outlet ports (SID3) was actuated once and thrice, the amount of drug release increased from 123.0 ± 6.99 μg to 357.3 ± 9.70 μg, respectively, which was reproducible over 30 days. When the SID3 was implanted in living animals for 30 days, plasma drug concentration was measured to be 92–146 ng ml−1 or 210–363 ng ml−1 when the device was actuated once or three consecutive times, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.