Abstract

The 2018 United Network for Organ Sharing allocation policy change has led to a significant increase in the use of mechanical circulatory support devices in patients listed for orthotopic heart transplantation. However, there has been a paucity of data regarding the newest generation Impella 5.5, which received FDA approval in 2019. The United Network for Organ Sharing registry was queried for all adults awaiting orthotopic heart transplantation who received Impella 5.5 support during their listing period. Waitlist, device, and early post-transplant outcomes were assessed. A total of 464 patients received Impella 5.5 support during their listing period with a median waitlist time of 19 days. Among them, 402 (87%) patients were ultimately transplanted, with 378 (81%) being directly bridged to transplant with the device. Waitlist death (7%) and clinical deterioration (5%) were the most common reasons for waitlist removal. Device complications and failure were uncommon (<5%). The most common post-transplant complication was acute kidney injury requiring dialysis (16%). Survival at 1-year post-transplant survival was 89.5%. Since its approval, the Impella 5.5 has been increasingly used as a bridge to transplant. This analysis demonstrates robust waitlist and post-transplant outcomes with minimal device-related and postoperative complications.