Abstract

In this work, an electrochemical immunosensor to determine patulin mycotoxin in apple juice is developed. It consists of a glassy carbon electrode modified with a dispersion of graphene oxide and polyclonal anti-patulin antibodies bound to their oxygenated groups. Patulin determination was performed by electrochemical impedance spectroscopy. The calibration curve is linear for a concentration range from 1 × 10−2 to 10 ng mL−1. The limit of detection and the concentration of patulin to obtain 50% inhibition (IC50) were 9.8 pg mL−1 and 360 pg mL−1, respectively. In addition, an acceptable accuracy, with a percentage coefficient of variation (%CV) close to 20%, and recovery percentage (%R) of 86% were found. The electrochemical immunosensor has the advantage of allowing the determination of patulin without any pre-treatment of the samples. The results obtained from the electroanalytical technique were compared with those obtained by HPLC.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.