Abstract

Using electrochemical impedance spectroscopy (EIS) the sensitive and specific detection of the antibiotic resistance gene mecA has been demonstrated. The gene sequence was obtained from clinical Staphylococcus aureus isolates. Initially a mecA specific probe was selected from hybridisation tests with a 3′ and 5′ version of a previously published probe sequence. When immobilised on a gold electrode in PNA form it was possible to detect hybridisation of mecA PCR product electrochemically at concentrations as low as 10nM. By incorporating an undecane-thiol and 1.8nm glycol spacer into the PNA probe it was possible to extend the limit of detection for mecA to 10 pM. Most published studies on EIS and nucleic acid detection report the use of short artificial DNA sequences or novel signal amplification schemes which improve sensitivity whereas this study reports the successful detection of long DNA fragments produced by PCR following extraction from clinical isolates. Finally, using screen printed electrodes the paper demonstrates hybridisation monitoring of mecA in an “on-line” assay format under ambient conditions which paves the way for rapid mecA detection in point of care scenarios.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.