Abstract

We report here a non-invasive, reversible method for interrogating single cells in a microfluidic flow-through system. Impedance spectroscopy of cells held at a micron-sized pore under negative pressure is demonstrated and used to determine the presence and viability of the captured cell. The cell capture pore is optimized for electrical response and mechanical interfacing to a cell using a deposited layer of parylene. Changes in the mechanical interface between the cell and the chip due to chemical exposure or environmental changes can also be assessed. Here, we monitored the change in adhesion/spreading of RAW264.7 macrophages in response to the immune stimulant lipopolysaccharide (LPS). This method enables selective, reversible, and quantitative long-term impedance measurements on single cells. The fully sealed electrofluidic assembly is compatible with long-term cell culturing, and could be modified to incorporate single cell lysis and subsequent intracellular separation and analysis.