Abstract

Thoracic impedance is modulated by events within the respiratory cycle, which represents a source of "noise" in impedance cardiography. Respiration itself, however, is a physiological rhythm of interest to psychophysiologists. We report here methods and validation for deriving impedance pneumographic measures of respiration from impedance cardiography signals, based on standard tetrapolar band electrodes. We recorded the change in impedance (delta Z), the first derivative of the change in impedance (dZ/dt), output from a strain-gauge respirometer, and criterion spirometry from eight healthy adults during rest, paced breathing, abdominal breathing, thoracic breathing, and a mental arithmetic task. Transfer function analyses revealed that a delta Zd signal (derived by integration of the dZ/dt signal) provided the best estimate of the criterion spirometric measure for all parameters (coherence, phase, and gain), accounting for almost 90% of the variance in respiratory waveform morphology. The results document the potential utility of impedance pneumography, as derived from standard impedance cardiography signals.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.