Abstract

In this double-blind, crossover study the authors have validated stroke volume determination by impedance cardiography against the pulsed Doppler echocardiographic method in elderly hypertensives. They found a good correlation between the stroke volume values obtained by the two methods over a range of values from 30 to 130 mL. The coefficient of linear regression was about .95 at each visit. The mean of the differences was -0.73 mL with a standard deviation of 8.46. Given that individual differences are normally distributed, the values corresponding to 2 standard deviations of the mean define a range covering 95% of the observed differences. From the distribution of the data around the mean plot it appears that, in comparison with pulsed Doppler, impedance cardiography tends to slightly underestimate stroke volumes of greater than 90 mL and to overestimate values of less than 50 mL. The results of this study indicate that impedance cardiography may represent a reliable alternative to pulsed Doppler echocardiography for the noninvasive estimation of cardiac output at rest in elderly patients.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.