Abstract
A conductometric sensor for potassium ions in solution is presented. Interdigitated, planar gold electrodes were coated with a potassium-selective polymer membrane composed of a poly(vinyl chloride) matrix with about 65 wt% of plasticiser and 2–5 wt% of a potassium-selective ionophore. The impedance of the membrane was measured, using the electrodes as a transducer, and related to the concentration of potassium in a sample solution in contact with the membrane. Sensitivity was optimised by varying the sensor components, and selectivity for potassium over sodium was also shown. The resulting devices are compact, miniature, robust sensors which, by means of impedance measurements, eliminate the need for a reference electrode. The sensor was tested for potassium concentration changes of 2 mM across the clinically relevant range of 2.7–18.7 mM.
Submitted Version (Free)
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.