Abstract
ABSTRACT.Mismatched input electrode impedances can lead to noisy EEG and evoked potential recordings even if the electrode impedances are low. This paper presents a quantitative analysis of the phenomenon and illustrative examples.At each amplifier input, the recording electrode impedance and the amplifier's input impedance form a voltage divider. When input electrode impedances are unequal, a common mode signal that is present equally at the two locations on the patient's body appears at different magnitudes at the two (inverting and noninverting) inputs of the amplifier; thus, part of the common mode signal is amplified as if it were a differential signal. The effective common mode rejection ratio (CMRR) of the amplifier is reduced to a value that can be expressed mathematically as a function of the amplifier's ideal CMRR, its input impedance, and the impedance difference between the two input electrodes.Balanced input electrode impedances are particularly important during intraoperative monitoring, whe...
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.