Abstract

This research aimed to examine the impact of NaOCl and Irritrol solutions with/without autoclave sterilisation on the cyclic fatigue resistance of ProTaper, WaveOne Gold and Reciproc NiTi files. Seventy files in every group were separated into seven subgroups (n = 10) according to irrigation solutions and sterilisation procedures. The mean and standard deviations of the time to fracture was calculated for each group, and the data statistical analysis was performed (P ≤ 0.05). Significantly higher cyclic fatigue resistance was found in Reciproc files than the other types of files. The differences between the control and other subgroups were found to be statistically insignificant (P > 0.05), except for subgroups 5 and 7 (P ≤ 0.05). Immersing NiTi files in NaOCl or Irritrol with sterilisation for three times considerably reduced the cyclic fatigue resistance of these files. Reciproc files exhibited the highest cyclic fatigue resistance in comparison with ProTaper and WaveOne Gold files.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.