Abstract
Removing the swab after collection can speed up diagnosis and improve the quality of laboratory procedures. This study investigates the impact of swab removal on SARS-CoV-2 detection in clinical specimens with a focus on high Cycle threshold (Ct) samples (Cts≥32). The method assessed pairs of SARS-CoV-2 samples mimicking combined throat and nose swabs and tested them on two real-time-PCR platforms; the Applied Biosystems 7500 and the Abbott Alinity. Swab removal did not significantly affect detection rates of SARS-CoV-2 samples with Ct values<32, regardless of the PCR platform. However, reduced reproducibility was seen at the endpoint limit of detection of the platforms, which meant that fewer samples with Ct values≥32 were detected in the swab removal group.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.