Abstract

BackgroundWhile 68Ga-PSMA PET-MRI might be superior to PET-CT with regard to soft tissue assessment in prostate cancer evaluation, it is also known to potentially introduce additional PET image artefacts. Therefore, the impact of PET acquisition duration and attenuation data on artefact occurrence, lesion detectability, and quantification was investigated.To this end, whole-body PET list mode data from 12 patients with prostate cancer were acquired 1 h after injection of 2 MBq/kg [68Ga]HBED-CC-PSMA on a hybrid PET-MRI system. List mode data were further transformed into data sets representing 300, 180, 90, and 30 s acquisition duration per bed position. Standard attenuation and scatter corrections were performed based on MRI-derived attenuation maps, complemented by emission-based attenuation data in areas not covered by MRI. A total of 288 image data sets were reconstructed with varying acquisition durations for emission and attenuation data with and without scatter and prompt gamma correction, and further analysed regarding image quality and diagnostic performance.ResultsDecreased PET acquisition durations resulted in a significantly increased incidence of halo artefacts around kidneys and bladder, decreased lesion detectability and lower SUV as well as markedly lower arm attenuation values: Halo artefacts were present in 5 out of 12 cases at 300-s duration, in 6 at 180 s, in 10 at 90 s, and in 11 cases at 30 s. Using attenuation data of the 300 s scans restored artefact occurrence to the original 300-s level. Prompt gamma correction only led to small improvements in terms of artefact occurrence and size. Of the 141 detected lesions in the 300-s images one lesion was not detected at 180 s, 28 at 90 s, and 64 at 30 s. Using the 300-s attenuation map decreased non-detectability of lesions to zero at 180 s, 9 at 90 s, and 52 at 30 s. Attenuation maps at 90 and 30 s demonstrated markedly lower mean arm attenuation values (0.002 cm-1) than those at 300 s (0.084 cm-1), and 180 s (0.062 cm-1).ConclusionsShort acquisition durations of less than 3 minutes per bed position result in unacceptable image artefacts and decreased diagnostic performance in current whole-body 68Ga-PSMA PET-MRI and should be avoided. Increased image noise and imperfections in generated attenuation maps were identified as a paramount cause for image degradation.

Highlights

  • While 68Ga-prostate specific membrane antigen (PSMA) positron emission tomography (PET)-Magnetic resonance imaging (MRI) might be superior to PET-Computed tomography (CT) with regard to soft tissue assessment in prostate cancer evaluation, it is known to potentially introduce additional PET image artefacts

  • When using the attenuation map derived from 300-s emission data for attenuation correction (AC) and scatter correction (SC), no such increase in halo artefacts with reduction of acquisition duration was observed, and halo scores remained on the level of 300-s emission data (Fig. 2)

  • Emission data obtained from the full 300-s data set reconstructed with attenuation maps of decreased scanning durations again resulted in increasing artefact numbers (Fig. 3), demonstrating the profound effect of maximum likelihood reconstruction of activity and attenuation (MLAA)-based arm completion on halo artefact appearance

Read more

Summary

Introduction

While 68Ga-PSMA PET-MRI might be superior to PET-CT with regard to soft tissue assessment in prostate cancer evaluation, it is known to potentially introduce additional PET image artefacts. The impact of PET acquisition duration and attenuation data on artefact occurrence, lesion detectability, and quantification was investigated. To this end, whole-body PET list mode data from 12 patients with prostate cancer were acquired 1 h after injection of 2 MBq/kg [68Ga]HBED-CC-PSMA on a hybrid PET-MRI system. A total of 288 image data sets were reconstructed with varying acquisition durations for emission and attenuation data with and without scatter and prompt gamma correction, and further analysed regarding image quality and diagnostic performance. A current study suggests a superiority of 68Ga-PSMA PET-MRI for the detection of prostate cancer lesions compared to PET-CT [9]

Objectives
Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.