Abstract

Ultraviolet UV absorbance measurements are subject to significant error using a standard spectrophotometer when particles or aggregates that scatter light are present. True UV absorbance for highly turbid waters should be measured using integrating sphere IS spectrophotometry that allows the collection of reflected and transmitted radiation simultaneously. This is especially important when the effects of scattering impact UV disinfection—such as with the presence of aggregates. The impact of light scattering of particle- aggregated microbes on UV disinfection was evaluated by comparing standard spectrophotometer and integrating sphere absorbance measurements for UV fluence determination. Spore-clay aggregates in simulated drinking waters and spore aggregates with natural particles from raw waters were induced by flocculation with alum. Coagulated systems significantly decreased the UV inactivation effectiveness compared to the noncoagulated system with the effects more pronounced for raw natural water. Absorbance measurement of suspensions and aggregates using standard spectrophotometry in the calculations of fluence resulted in overdosing whereas the use of IS spectroscopy did not. The results demonstrated that aggregation protected spores from UV disinfection, and that use of proper absorbance measurement techniques, accounting for particle scattering, is essential for correct interpretation of the results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.