Abstract

Technological advancements in newer-generation catheterisation laboratories may reduce patient and occupational radiation exposure. We compared fluoroscopy time and dose-area product (DAP) between a Philips Allura X-PER FD20 and Siemens Artis Zeego Hybrid systems for 47 single-vessel percutaneous coronary interventions (PCI) and 35 transcatheter aortic valve implantations (21 Corevalve, 14 Edwards Sapien TAVI) using the FD20, versus 30 PCI and 28 TAVI (15 Corevalve, 13 Sapien) with the Zeego over a 24-month period. Multivariate analysis revealed that, adjusting for patient weight and fluoroscopy time, DAP (median, interquartile range) was 26% lower for PCI with the Zeego than the FD20 [55.6 (27.0-91.5) vs 77.6 (51.2-129.1) Gy.cm(2), P=0.03)] and using tomographic imaging with the Zeego did not increase DAP for TAVI procedures [98.1 (65.9-136.6) vs 112.4 (64.9-156.2) Gy.cm(2) (P=NS). Although fluoroscopy times were longer for TAVI procedures than PCI with both systems (23.5-24.4 vs 7.3-9.2mins, p<0.0001), there was a significant difference in DAP between PCI and combined TAVI with the Zeego (55.6 vs 112.4Gy.cm(2), P<0.006) but not with the FD20 (77.6 vs 98.1Gy.cm(2), P=NS). Specific dose-reducing features of the new-generation system reduced DAP more for PCI than TAVI, as valve replacement procedures use additional cine-acquisition not necessary for PCI.