Abstract

To assess the impact of the width of multileaf collimator (MLC) on dose distributions on HyperArc fractionated stereotactic irradiation for multiple (5-10) brain metastases. Twenty-one HyperArc (HA) plans were generated using the high definition (HD) MLC (2.5 mm) to deliver 30-35 Gy in 3-5 fractions (HA-HD). The HyperArc plans using Millennium (ML) MLC (5 mm) were retrospectively generated (HA-ML) using the same planning parameters with HA-HD. Dosimetric parameters between the planning target volume (PTV) and organs at risk (OARs) were compared. The conformity index was significantly higher (p<0.0001) in the HA-HD plans (0.95±0.04) than that in the HA-ML plans (0.92±0.06). The HA-HD provided significantly lower (p<0.0001) gradient index (5.6±2.5) than HA-ML (6.2±3.5). For the brainstem and retina (right), a statistically significant difference (p<0.05) was observed between the HA-HD (12.8±10.9 and 2.8±1.7 Gy, for brainstem and retina, respectively) and HA-ML (13.6±11.1 and 3.0±1.8 Gy) plans. For the brain tissue, the HA-HD plans statistically significantly reduced dosimetric parameters (p<0.0001) in all evaluated dose range (V6Gy-V28Gy). The narrower MLC provided significantly higher conformity, steeper dose gradient, and better normal tissue sparing.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.