Abstract

The radiation dose during digital subtraction angiography (DSA) examination is determined on the basis of many factors. The aim of this study was to evaluate the dose-area product, air kerma, and fluoroscopy time on monoplane and biplane DSA of the cerebral arteries. Our results show that biplane angiography offered lower DAP, cumulative air kerma at the reference point and fluoroscopy time despite comparable procedure time.. Further research in this area and personnel training warrant the improvement of the radiosafety of angiographic interventional procedures and further development of angiographic technology.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.