Abstract

Magnetic resonance (MR)-linac delivery is expected to improve organ at risk (OAR) sparing. In this study, OAR doses were compared for online adaptive MR-linac treatments and conventional cone beam computed tomography (CBCT)-linac radiotherapy, taking into account differences in clinical workflows, especially longer session times for MR-linac delivery. For 25 patients with pelvic/abdominal lymph node oligometastases, OAR doses were calculated for clinical pre-treatment and daily optimized 1.5T MR-linac treatment plans (5×7Gy) and compared with simulated CBCT-linac plans for the pre-treatment and online anatomical situation. Bowelbag and duodenum were re-contoured on MR-imaging acquired before, during and after each treatment session. OAR hard constraint violations, D0.5cc and D10cc values were evaluated, focusing on bowelbag and duodenum. Overall, hard constraints for all OAR were violated less often in daily online MR-linac treatment plans compared with CBCT-linac: in 5% versus 22% of fractions, respectively. D0.5cc and D10cc values did not differ significantly. When taking treatment duration and intrafraction motion into account, estimated delivered doses to bowelbag and duodenum were lower with CBCT-linac if identical planning target volume (PTV) margins were used for both modalities. When reduced PTV margins were achievable with MR-linac treatment, bowelbag doses were lower compared with CBCT-linac. Compared with CBCT-linac treatments, the online adaptive MR-linac approach resulted in fewer hard planning constraint violations compared with single-plan CBCT-linac delivery. With respect to other bowelbag/duodenum dose-volume parameters, the longer duration of MR-linac treatment sessions negatively impacts the potential dosimetric benefit of daily adaptive treatment planning.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call