Abstract

Evaluating cardiac dose during total body irradiation (TBI) is of increasing interest. A three-dimensional beam model for TBI was commissioned and lung shielding was simulated in a treatment planning system with the cardiac silhouette partially blocked and unblocked. When blocked, the median heart dose decreased by 6% (IQR=6%) and the median cardiac V12Gy decreased by 27% (IQR=17%). The median left anterior descending artery dose decreased 20% (IQR=12%) for blocked cases. Because using partial heart shielding may result in considerable changes in dose to cardiac structures, TBI protocols should explicitly consider lung block design parameters and their potential effects.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.