Abstract

Ten subjects underwent treadmill exercise at 5.6 km/h over one hour while wearing each of three identical appearing, cup-shaped, prototype filtering facepiece respirators that differed only in their filter resistances (3 mm, 6 mm, and 9 mm H2O pressure drop). There were no statistically significant differences between filtering facepiece respirators with respect to impact on physiological parameters (i.e., heart rate, respiratory rate, oxygen saturation, transcutaneous carbon dioxide levels, tympanic membrane temperature), pulmonary function variables (i.e., tidal volume, respiratory rate, volume of carbon dioxide production, oxygen consumption, or ventilation), and subjective ratings (i.e., exertion, thermal comfort, inspiratory effort, expiratory effort and overall breathing comfort). The nominal filter resistances of the prototype filtering facepiece respirators correspond to airflow resistances ranging from 2.1 - 6.6 mm H2O/L/s which are less than, or minimally equivalent to, previously reported values for the normal threshold for detection of inspiratory breathing resistance (6 - 7.6 mm H2O/L/sec). Therefore, filtering facepiece respirators with filter resistances at, or below, this level may not impact the wearer differently physiologically or subjectively from those with filter resistances only slightly above this threshold at low-moderate work rates over one hour.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.