Abstract
This study compared results from plating urine specimens with the BD InoqulA instrument using a 10-μl inoculum with results from cultures plated manually with a 1-μl loop for comparable 2-month periods. The positivity rates, turnaround times for positive cultures, and BD Phoenix identification and antimicrobial susceptibility test results were comparable for both time periods. We experienced no problems with culture interpretation as the result of moving to the 10-μl inoculum.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.