Abstract
The accurate measurement of thoron activity concentration is an important issue in both thoron exposure evaluation and in reducing its influence on radon measurement. For radon monitors based on electrostatic collection technique and an alpha spectrometry analyser, air humidity and sampling flowrate are key factors influencing the sensitivity of thoron activity concentration measurement. For the purpose of improving thoron measurement sensitivity and stability, theoretical derivation and experimental studies were systemically performed in this study. The results show that thoron measurement sensitivity decreases as a negative exponential function with absolute humidity increasing, and the sensitivity of thoron is much lower than that of radon under the same conditions, which is mainly caused by the small value of the concentration ratio of thoron inside to outside of the chamber. When the air exchange rate of the measurement chamber (sampling flowrate/inner volume) increases, the measurement sensitivity of thoron gas first increases rapidly and then decreases slowly after reaching its maximum at the air exchange rate of 0.24 s−1. In practice, in the normal air exchange rate range (for example <0.05 s−1), increasing the sampling flowrate could greatly improve the thoron measurement sensitivity, which consequently suggests an effective way to update thoron measurement under the present conditions of the monitor.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.