Abstract

Background: Trans-nasal pulmonary aerosol delivery using high flow nasal cannula (HFNC) devices is described with the administration of high gas flows exceeding patient inspiratory flow (HF) and with lower flows (LF). The aim of this pilot clinical trial was to compare deposition and distribution of radiolabeled aerosol via nasal cannula in healthy adults across three rates of gas flow delivered with active heated humidification, and to further identify the impact of aerosol administration without heated humidity. Methods: Twenty-three (23) healthy adults (16F) were randomized to receive aerosol with active heated humidification or unheated oxygen at gas flows of 10 L/min (n = 8), 30 L/min (n = 7), or 50 L/min (n = 8). Diethylenetriaminepentaacetic acid labeled with 1 millicurie (37 MBq) of Technetium-99m (DTPA-Tc99m) was mixed with NaCl to a fill volume of 1 mL, and administered via mesh nebulizer placed at the inlet of the humidifier. Radioactivity counts were performed using a gamma camera and the regions of interest (ROIs) were delimited with counts from the lungs, upper airways, stomach, nebulizer, circuit, and expiratory filter. A mass balance was calculated and each compartment was expressed as a percentage of the total. Results: Lung deposition (mean ± SD) with heated humidified gas was greater at 10 L/min than 30 L/min or 50 L/min (17.2 ± 6.8%, 5.71 ± 2.04%, and 3.46 ± 1.24%, respectively; p = 0.0001). Using unheated carrier gas, a lung dose of aerosol was similar to the active heated humidification condition at 10 L/min, but greater at 30 and 50 L/min (p = 0.011). Administered gas flow and lung deposition were negatively correlated (r = −0.880, p < 0.001). Conclusions: Both flow and active heated humidity inversely impact aerosol delivery through HFNC. Nevertheless, aerosol administration across the range of commonly used flows can provide measurable levels of lung deposition in healthy adult subjects (NCT 02519465).

Highlights

  • Oxygen administration via high flow nasal cannula (HFNC) supports acute and critically ill patients with respiratory failure [1,2,3,4,5]

  • Trans-nasal pulmonary administration of medical aerosols using HFNC devices has been reported with gas flow exceeding patient inspiratory demand and gas delivery lower than patient inspiratory flow (LF) [10]

  • mass median aerodynamic diameter (MMAD) of aerosol exiting the cannula at 10, 30, and 50 L/min was greater with active heated humidification than unheated gas (2.29 ± 0.22 μm vs. 1.29 ± 0.22 μm; p = 0.038) with no trend in particle size change across flows

Read more

Summary

Introduction

Oxygen administration via high flow nasal cannula (HFNC) supports acute and critically ill patients with respiratory failure [1,2,3,4,5]. Trans-nasal pulmonary aerosol delivery using high flow nasal cannula (HFNC) devices is described with the administration of high gas flows exceeding patient inspiratory flow (HF) and with lower flows (LF). The aim of this pilot clinical trial was to compare deposition and distribution of radiolabeled aerosol via nasal cannula in healthy adults across three rates of gas flow delivered with active heated humidification, and to further identify the impact of aerosol administration without heated humidity. Methods: Twenty-three (23) healthy adults (16F) were randomized to receive aerosol with active heated humidification or unheated oxygen at gas flows of 10 L/min (n = 8), 30 L/min (n = 7), or 50 L/min (n = 8). Aerosol administration across the range of commonly used flows can provide measurable levels of lung deposition in healthy adult subjects (NCT 02519465)

Objectives
Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.