Abstract

BackgroundWe sought to compare diffusion tensor imaging (DTI) parameters in vivo and ex vivo in the brain and to explore the effects of radiofrequency coil and number of excitations on ex vivo DTI parameters.MethodsSix Sprague–Dawley rat brains were used to obtain in vivo and ex vivo DTI maps with different coils and number of excitations. DTI parameters of white matter and grey matter including diffusivities, fractional anisotropy, and other dimensionless ratios (λ2/λ1, λ3/λ1, and λ2/λ3) were obtained from reconstruction maps. Comparisons of ex vivo signal-to-noise ratio with different coils and number of excitations were conducted.ResultsDiffusivities decreased significantly after fixation in all the selected white matter and grey matter regions of interest (all at p < 0.001). The diffusivities in white matter integrity decreased more than in grey matter integrity after fixation (all at p < 0.001). The ratio of λ2/λ3 in the major brain structures changed after fixation (most at p < 0.05). There were differences in major ex vivo brain structures in DTI parameters and signal-to-noise ratio between surface coil and volume coil, and between one and four excitations (most at p < 0.05).ConclusionThe impact of fixation, coil, and number of excitations on DTI parameters should be taken into consideration in clinical and experimental studies at 7.0 T.

Highlights

  • We sought to compare diffusion tensor imaging (DTI) parameters in vivo and ex vivo in the brain and to explore the effects of radiofrequency coil and number of excitations on ex vivo DTI parameters

  • Previous studies have compared in vivo and ex vivo characteristics of DTI, finding that the diffusion decreased after fixation [12–14]

  • The results for the radial diffusivity (RD), ratios of λ2/λ1 (R1), and λ3/λ1 ratio (R2) were opposite to those for the fractional anisotropy (FA), and the Axial diffusivity (AD) in white matter (WM) integrity and grey matter (GM) integrity was similar, suggesting that the RD constitutes the main difference in diffusion between WM integrity and GM integrity

Read more

Summary

Introduction

We sought to compare diffusion tensor imaging (DTI) parameters in vivo and ex vivo in the brain and to explore the effects of radiofrequency coil and number of excitations on ex vivo DTI parameters. Methods: Six Sprague–Dawley rat brains were used to obtain in vivo and ex vivo DTI maps with different coils and number of excitations. DTI parameters of white matter and grey matter including diffusivities, fractional anisotropy, and other dimensionless ratios (λ2/λ1, λ3/λ1, and λ2/λ3) were obtained from reconstruction maps. Comparisons of ex vivo signal-to-noise ratio with different coils and number of excitations were conducted. Ex vivo DTI is widely used to acquire high-quality images with high resolution in clinical and basic studies with long-time scanning [10–12].

Methods
Results
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call