Abstract

Near-infrared diffuse optical tomography (NIR-DOT) is an emerging technology that offers hemoglobin based, functional imaging tumor biomarkers for breast cancer management. The most promising clinical translation opportunities are in the differential diagnosis of malignant vs. benign lesions, and in early response assessment and guidance for neoadjuvant chemotherapy. Accurate quantification of the tissue oxy- and deoxy-hemoglobin concentration across the field of view, as well as repeatability during longitudinal imaging in the context of therapy guidance, are essential for the successful translation of NIR-DOT to clinical practice. The ill-posed and ill-condition nature of the DOT inverse problem makes this technique particularly susceptible to model errors that may occur, for example, when the experimental conditions do not fully match the assumptions built into the image reconstruction process. To evaluate the susceptibility of DOT images to experimental errors that might be encountered in practice for a parallel-plate NIR-DOT system, we simulated 7 different types of errors, each with a range of magnitudes. We generated simulated data by using digital breast phantoms derived from five actual mammograms of healthy female volunteers, to which we added a 1-cm tumor. After applying each of the experimental error types and magnitudes to the simulated measurements, we reconstructed optical images with and without structural prior guidance and assessed the overall error in the total hemoglobin concentrations (HbT) and in the HbT contrast between the lesion and surrounding area vs. the best-case scenarios. It is found that slight in-plane probe misalignment and plate rotation did not result in large quantification errors. However, any out-of-plane probe tilting could result in significant deterioration in lesion contrast. Among the error types investigated in this work, optical images were the least likely to be impacted by breast shape inaccuracies but suffered the largest deterioration due to cross-talk between signal channels. However, errors in optical images could be effectively controlled when experimental parameters were properly estimated during data acquisition and accounted for in the image processing procedure. Finally, optical images recovered using structural priors were, in general, less susceptible to experimental errors; however, lesion contrasts were more sensitive to errors when tumor locations were used as a priori info. Findings in this simulation study can provide guidelines for system design and operation in optical breast imaging studies.

Highlights

  • Near-infrared diffuse optical tomography (NIR-DOT) is emerging as a useful imaging tool for breast cancer management [1,2,3,4,5,6]

  • Similar trends in μs'690 images are expected since μs' at 690 nm and 830 nm are calculated from the same amplitude and power components of a power law fit obtained from the Redbird reconstruction of reduced scattering. 3.3 Tolerance to experimental errors To enable direct comparisons of the influence of experimental errors on all phantom cases using various reconstruction methods, two metrics were introduced to determine the tolerance of recovered optical properties to errors, i.e., the maximum extent of each experimental error at which reconstructed optical images still yield acceptable quality/accuracy

  • While slight in-plane probe misalignment and plate rotation did not result in large quantification errors, out-of-plane probe tilting that is not accounted for can result in significant deterioration in lesion contrast

Read more

Summary

Introduction

Near-infrared diffuse optical tomography (NIR-DOT) is emerging as a useful imaging tool for breast cancer management [1,2,3,4,5,6]. Optical imaging has been demonstrated to offer early sensitivity to tissue changes predictive of therapy outcome in patients undergoing neoadjuvant chemotherapy [7,8,9,10,11,12,13,14]. Factors relating to patient positioning repeatability, variation in optode coupling and breast shape, as well as accuracy in generating the finite element mesh used for reconstruction can negatively impact the sensitivity of optical tissue monitoring. Methods that are designed to define and ensure quantification accuracy, precision, and reproducibility, especially over the long time frame of therapy monitoring studies, are essential for the clinical translation of NIR-DOT [17]

Objectives
Methods
Results
Discussion
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call