Abstract

To assess the effect of cochlear implant (CI) insertion depth and surgical technique on intracochlear trauma. Twenty-one fresh human temporal bones were implanted with CI electrodes and underwent histologic processing and evaluation. Specimens were grouped into 3 categories: 1) soft implantation technique and standard electrode; 2) soft implantation technique and flexible prototype array; 3) forceful implantations and standard electrode. Based on the grading system (1 to 4), 2 numeric values were calculated indicating the overall severity of cochlear damage (trauma indices). Mean trauma index values were 13.8, 36.3, and 59.2 for group 1, 2, and 3, respectively. Differences in cochlear trauma (trauma index) were nonsignificant between specimens in groups 1 and 2 but were significant between groups 1 and 3. This study gives evidence that intracochlear trauma increases with deep insertions. Thus, in cases where cochlear integrity might be important, limited insertions should be achieved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.