Abstract
We evaluated the effect of different xSPECT Bone acquisition methods on image quality and quantitative accuracy. A mixed bone-equivalent solution set to 250 HU and 99mTc was filled with the SIM2 Bone phantom. The xSPECT Bone acquisitions were performed with the step and shoot (SS) mode and two continuous modes (continuousview, continuoustotal) of the times per view, and the total acquisition times were the same as the SS mode. The radioactive concentration ratio of normal bone and tumor was set to 1 : 6. The reconstructed images were evaluated using uniformity, contrast, spatial resolution, recovery coefficient (RC) of the tumor part, and detectability. The detectability score (DS), an overall evaluation of detectability, was calculated from the results of uniformity and contrast. Uniformity of SS, continuousview, and continuoustotal methods was 29.4%, 17.4%, and 11.1%, respectively. Spatial resolution, RC, and detectability were similar among the SS mode and two continuous modes; however, the continuous modes showed good uniformity and contrast compared with the SS mode. In addition, the detectability of the continuous mode was equal to or superior to that of the SS mode.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.