There is no established hemostasis method or protocol for the transdistal radial approach. Therefore, this study aimed to determine whether “the PreludeSYNC DISTAL” radial compression device (PSD; Merit Medical Systems, Inc., South Jordan, UT) can effectively prevent distal radial artery (dRA) occlusion following catheterization procedures. This retrospective study analyzed patients who underwent hemostasis using the PSD from January 1, 2019, to March 31, 2019. The primary endpoint was occlusion or excessive stenosis of the radial artery (RA) 1 month after catheterization. Pulsatile blood flow and vessel diameters of the dRA and forearm RA (fRA) were measured using vascular ultrasound before and 1 month after catheterization to determine arterial damage. Secondary endpoints were achievement of hemostasis, bleeding, hematoma, aneurysm, neurological abnormality, and functional disturbance of the fingers or hand. Fifty patients (mean age, 70.9 ± 10.7 years; male, 72.0%) were enrolled in this study. Complete hemostasis was achieved in all cases. Total hemostasis time was 161 ± 45 min. No procedure-associated complications were noted. Pulsations of the dRA and fRA were maintained at 1 month. No functional disturbance or neurological abnormality was observed. Vessel diameters of the dRA and fRA were not significantly different before and 1 month after catheterization. No dissection, pseudoaneurysm, or occlusion/stenosis was observed on ultrasound. Distal radial access with a unique device and protocol effectively achieved hemostasis and prevented injury and occlusion of the dRA and fRA.

Highlights

  • Percutaneous coronary intervention (PCI) is an established treatment for ischemic heart disease [1,2,3]

  • BARCBleeding Academic Research Consortium, distal RA (dRA) distal radial artery, forearm RA (fRA) forearm radial artery. This is the first report of the dRA condition observed using ultrasound and a patent hemostasis protocol with a dedicated hemostasis device after the transdistal radial approach to coronary catheterization

  • Using the PreludeSYNC DISTAL radial compression device (PSD) and a simple hemostatic protocol, blood flow in both the dRA and fRA without damage, as observed using ultrasound was maintained 1 month after catheterization and complete hemostasis without complications was achieved in all cases

Read more

Summary

Introduction

Percutaneous coronary intervention (PCI) is an established treatment for ischemic heart disease [1,2,3]. A factor that affects the prognosis of PCI is the site used during the approach. A new approach involving coronary catheterization via the distal RA (dRA) at the anatomical snuffbox has been reported [9]. An advantage of this method is its shorter hemostasis time than conventional TRA [10, 11]. No established hemostasis method or protocol for the transdistal radial approach and no suitable method of hemostasis using dRA access and a dedicated device have been reported.

Objectives
Methods
Results
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call