Abstract

AbstractAim:To compare dose to target and organs at risk (OARs) in conventional plan (2D) versus computed tomography (CT)-based three-dimensional (3D) plan in vaginal cuff brachytherapy (VBT) and to compare the effect of bladder distension on target and OARs dosimetry.Materials and methods:Post-hysterectomy patients with an indication for VBT were included in the study. All patients underwent planning CT scans with a full bladder and an empty bladder protocol. For each CT, two plans were generated—one library-based 2D plan and another CT-based 3D plan. Dosimetric parameters were recorded for clinical target volume (CTV) and OARs.Results:A total of 92 observations were made from data collected from 46 patients. Difference between CTV dose in terms of 2D and 3D plans were not statistically significant for CTV (p = 0·11). Significant reduction in D0·1cc, D1cc and D2cc dose parameters were observed in bladder, rectum, sigmoid and bowel doses with the 3D plan (p < 0·001). Bladder distension showed a 20% reduction in dose for bowel (p < 0·001). Bladder distension also showed a 6·12% (p = 0·047) increase in D2cc, but there was a significant reduction in the mean dose to the bladder.Conclusion:Our study demonstrates the dosimetric benefits with 3D CT-based planning for VBT over 2D-based conventional planning and benefit of bladder distension in the reduction of bowel dose without compromising dose to the target volume.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call