Abstract

This study aimed to evaluate the collimation practices and quantify the overexposure due to extensive X-ray field area. The study was carried out in four digital X-ray units (including one paediatric X-ray unit). A total of 749 X-ray projections (555 adult and 194 paediatric) were evaluated. In adult X-ray units, the radiation field size was two times larger than the electronically collimated field. In the paediatric unit, the radiation field was 3.7 times larger than the electronic collimated field. The average additional entrance surface dose due to the excess radiation field used in a paediatric X-ray unit varied between 9.3 (2.5%) and 201.4 $\mu $Gy (10.9%). Therefore, proper pre-patient collimation should be applied whenever feasible, which reduces the patient radiation dose considerably.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.