Abstract

PurposeThis study is aimed to evaluate the impact of catheter contact angle on lesion formation and durability of pulmonary vein isolation (PVI).MethodsBoth in vitro experiment and retrospective observational study were conducted. For in vitro experiment, radiofrequency lesions were created on explanted swine hearts in three different catheter contact angles (0°, 45°, and 90°). In the retrospective observational study, we assessed patients who had undergone repeat catheter ablation due to atrial fibrillation recurrence after initial PVI. When pulmonary vein (PV) reconnection was observed, we analyzed the previous ablation points within and without the gap area. The gap areas were where ablation had changed the PV activation sequence or eliminated the PV potential in the repeat session.ResultsIn the in vitro experiment, lesion width was the smallest (5.3 ± 0.4 mm) in perpendicular contact compared to 0° (vs 5.8 ± 0.5 mm, p = 0.040) and 45° (vs 6.4 ± 0.4 mm, p < 0.001). In the retrospective observational study, we assessed 666 tags of 16 patients with PV reconnections, and 60 tags were in the gap area. Tags in the gap area had longer interlesion distance (odds ratio [OR] 1.49, p < 0.001), greater contact force variability (OR 1.03, p = 0.008), and higher rate of perpendicular contact (OR 3.26, p < 0.001) on multivariate analysis.ConclusionsPerpendicular contact was associated with a smaller lesion and higher rate of PV reconnection.