Abstract

Infusion catheters, when used in combination with balloons, are subjected to pressure created by inflation of the balloon. The compression can reduce the catheter flow area and cause elevated shear stresses in the fluid. A model and experiments were developed with a range of applied balloon pressures to investigate whether such situations may cause cell lysis during stem-cell infusion through single-lumen catheters. It was found that for balloon inflation pressures in excess of ~7 atm, it is possible for cell injury to occur, although the critical pressure depends on the fluid viscosity. The study was then applied to a multi-lumen catheter which was designed to resist compression. That device was able to handle very elevated balloon pressures and flow rates before cell lysis became a concern.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.