Abstract

To assess whether asymmetric stent expansion affects suppression of neointimal hyperplasia after sirolimus-eluting stent implantation, 64 patients in the SIRolImUS-coated Bx Velocity stent trial who underwent single 18-mm stent implantation and 3-dimensional intravascular ultrasonography at 8-month follow-up were enrolled. To assess the longitudinal stent asymmetric expansion, 2 cross sections with a maximal/minimal stent area were chosen in each patient. To assess for tomographic stent asymmetric expansion, stent eccentricity was determined by dividing the minimum stent diameter by the maximum stent diameter. At the 2 cross sections with a maximal/minimal stent area, a sirolimus-eluting stent reduced neointimal hyperplasia significantly with no interaction between the treatment and stent areas. A sirolimus-eluting stent also significantly reduced neointimal hyperplasia in the concentric and eccentric stent groups.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.