Abstract

This study compares the mean glandular dose (MGD) across 2D, 3D projection and Contrast-Enhanced Digital Mammography (CEDM) mammographic techniques. The important metadata were extracted from the digital mammography console. 650 subjects were clustered based on projections, age and CBT. The MGD of 2D, 3D, and CEDM was positively correlated with CBT but inversely correlated with the age factor. This study indicate MGD of CEDM was 16% and 22% lower compared to 2D and 3D techniques, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.