Abstract

Bolak Eldakror Hospital is a secondary care governmental hospital in Giza, Egypt. the Gastrointestinal endoscopy unit was founded in 1999. A written protocol for endoscope decontamination based on international standards was implemented and an ongoing quality assurance program was established in 2003 when we commenced bacteriological monitoring of endoscopes. Pseudomonas aeruginosa was found in a specimen obtained from a colonoscope. the unit guidelines were reviewed. the cause of contamination was a breach in reprocessing technique. A nurse admitted that she did not remove or clean the cap in the distal tip of the colonoscope at the end of a long busy day. We reprocessed and sampled the colonoscope. the new specimen yielded no growth. the staffing system was changed to secure a balance between the number of cases and work hours. A disinfection logbook and reprocessing checklists were provided for the nurses to record decontamination procedures.