Abstract
Significant deviations between published results have been reported measuring the effective thermal conductivity of nanofluids with the transient hot-wire method (THWM). This may be attributed to a poor selection of the temperature data range, which should meet the following conditions. The start time should be chosen after the conductive heat flux delay time, while the end time should be selected before a crossover point when natural convection becomes significant. Considering an EG-based 1.06 vol.% ZnO nanofluid, the thermal conductivity was measured to increase by 5.4% over that of the base fluid.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.