Abstract

Biosensors based on laccase immobilized on microparticles of chitosan crosslinked with tripolyphosphate (biosensor I) and glyoxal (biosensor II) obtained by spray drying for the determinations of rutin in pharmaceutical formulations were developed. Under optimized operational conditions (pH 4.0, frequency of 30 Hz, pulse amplitude of 40 mV and scan increment of 2.0 mV) two analytical curves were obtained for both biosensors showing a detection limit of 6.2x10-8 mol L-1 for biosensor (I) and 2.0x10-8 mol L-1 for biosensor (II). The recovery of rutin from pharmaceutical sample ranged from 90.7 to 105.0% and the lifetime of these biosensors were 4 months (at least 400 determinations).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.