Abstract
The purpose of this research is to demonstrate a new design for a cortisol immunosensor for the noninvasive and quantitative analysis of salivary cortisol. We propose a cortisol immunosensor with a fluid control mechanism which has both a vertical flow and a lateral flow. The detected current resulting from a competitive reaction between the sample cortisol and a glucose oxidase (GOD)-labeled cortisol conjugate was found to be inversely related to the concentration of cortisol in the sample solution. A calibration curve using the relative detected current showed a R2=0.98 and CV=14% for a range of standard cortisol solutions corresponding to the concentrations of native salivary cortisol (0.1–10ng/ml). The measurement could be accomplished within 35min and the cortisol immunosensor could be reused. These results show promise for realizing an on-site and easy-to-use biosensor for cortisol. Used for evaluation of human salivary cortisol levels, the cortisol immunosensor measurement corresponded closely with commercially available ELISA method (R2=0.92). Our results indicate the promise of the new cortisol immunosensor for noninvasive, point of care measurement of human salivary cortisol levels.