Abstract

Recently, the N-terminal fragment of proatrial natriuretic peptide (N-terminal proANP) has been proposed as a marker of chronic congestive heart failure. In this study, we established a two-step immunoradiometric assay using monoclonal antibodies and synthetic N-terminal proANP (1-67) as a standard. It allows us to measure plasma N-terminal proANP in only 4 h without prior extraction. The detection limit of this assay was 15 pmol/L for a 100 microL sample of plasma. Within-run CVs ranged from 1.7% to 2.9% and between-run CVs ranged from 4.2% to 5.1%. The dilution curves of plasma samples showed good linearity and analytical recovery was 89-104%. The mean (+/-SD) N-terminal proANP in plasma of 33 healthy subjects was 188 (+/-71) pmol/L and 1030 (+/-411) pmol/L in 25 patients with heart failure. Our immunoradiometric assay is rapid and precise enough for routine determination of N-terminal proANP in human plasma.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.