Abstract

This study evaluated the extended use of a haematology analyser (Abbott Cell-Dyn CD4000) for the immunofluorescent enumeration of foeto-maternal haemorrhage (FMH) with fluorescein isothiocyanate-labelled monoclonal anti-RhD. Method performance was assessed with artificial FMH standards, and a series of 44 clinical samples. Within run precision was <15% (coefficient of variation, CV) for FMH volumes of 3 ml and above, 18.8% at an FMH volume of 2 ml and 31.7% at an FMH volume of 1 ml. Linearity analysis showed excellent agreement (observed FMH% = 0.98x expected FMH% + 0.02), and a close relationship (R(2) = 0.99) between observed and expected FMH percentages. The lower limit of quantification of the CD4000 (SRP-Ret) method with a maximum CV of 15% was 1.6 ml, and the limit of detection was <1 ml. Parallel Kleihauer-Betke test (KBT) assessments of FMH standards showed an overall trend for higher KBT values (observed = 1.25x expected - 0.38). At an FMH level of 4 ml, KBT observer estimates ranged from 0.57 to 11.94 ml with a mean inter-observer CV of 63%. For 44 clinical samples, there was decision point agreement between KBT and SRP-Ret results for 42 samples with an FMH of <2 ml. Analysis in the low FMH range (<1 ml) showed that small volume foetal leaks could be detected with the SRP-Ret method in most of 23 samples with negative KBT results. CD4000 SRP-Ret method performance for FMH determination was similar to that reported for flow cytometry.