Abstract

Abstract Two different immunoassay (IA) formats for pesticide residue analysis, a field format (dipstick IA) and an automated version (flow injection immunoanalysis, FIIA), are described with regard to method development and validation performed by measuring water, soil, and liquid food samples and comparing the results with the respective GC data obtained. As representative analyte a pesticide metabolite, deethylatrazine as the major atrazine metabolite, was selected. The measuring range for deethylatrazine was from 0.1–10 μg/L with the dipstick IA and from 0.01–10 μg/L with the FIIA. With both formats, water and liquid food samples can be measured directly without the need for any prior enrichment or clean-up steps. In the case of the soil samples, two extraction steps can be omitted as compared to extract preparation for GC analysis which usually takes further 3–4 h. A close correspondence was found between the results of the two immunoassay formats and GC measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.