Abstract
Summary We performed an immunoassay analysis using the Quartz Crystal Microbalance (QCM) biosensor to detect the specific binding reaction of the (Human IgG1)(Anti-Human IgG1) protein pair. Both experimental and computational were used to study biomolecular binding reactions in microfluidic channels. We discussed the unsteady convective diffusion in the transportation tube and found that the distribution of the analyte concentration in the tube is strongly affected by the flow field. Due to this, large discrepancies between the simulations and experimental results were observed. We show that the conventional assumption of the uniform and steady analyte concentration used in the computational procedure is not accurate enough. In addition, we show that the commonly used procedure in kinetic analysis for estimating binding rate constants from the experimental data would underestimate these rate constants due to neglected diffusion processes from the inlet to the reaction surface. A modified procedure is proposed to supplement the kinetic analysis, thus yielding better consistency with experiments.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.