Abstract

This work describes the fundamental ability of a commercial ELISA to determine acetamiprid and the application of the ELISA to residue analysis in fruit and vegetable samples. The ELISA exhibited satisfactory sensitivity (I (50) 0.6 ng/g; limit of detection 0.053 ng/g) and a high selectivity for acetamiprid versus other neonicotinoid analogs (thiacloprid amide). Methanol, which influenced the sensitivity of the ELISA the least, was selected as the extractant for the ELISA analysis. Simple dilution of sample extracts with water eliminated matrix interferences. Average recoveries from the acetamiprid-spiked agricultural samples were >95% using a simple extraction method. Analytical results obtained from the ELISA were comparable to those obtained from the reference HPLC method (r>0.99). The ELISA applied to the residue analysis of acetamiprid in agricultural products is a rapid, simple, and cost-effective method, and could be successfully applied to the detection of acetamiprid before the distribution of produce.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.