Abstract

A highly selective and sensitive method was developed for the simultaneous determination of four β-agonists (clenbuterol, salbutamol, ractopamine and terbutaline) in beef by immunoaffinity chromatography purification coupled to ultra-high-performance LC-MS/MS. The MS/MS conditions, ultra-high-performance LC mobile phase, injection solution, sample purification process and matrix effect were studied to optimise the operation conditions. The limits of detection (LODs) of the instrument for the studied β-agonists ranged from 0.20 to 0.25 µg l−1, and the LODs of the method for the studied β-agonists ranged from 0.20 to 3.00 µg kg−1 for beef. Calibration curves were constructed using a standard solution diluted with blank beef matrix. The linear ranges of the calibration curves ranged from 5 to 100 µg kg−1 and the coefficients of determination were >0.9942 (n = 10) for all four β-agonists. Samples spiked at 5, 10 and 50 µg kg−1 showed recoveries >72% and RSDs <6.6%. The method is suitable for the simultaneous detection of four β-agonists at trace levels in beef.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.