Abstract

An immune complex transfer two-site chemiluminescent immunoassay (CLIA) for salmon growth hormone (GH) was developed to measure serum GH in alevin chum salmon (Oncorhynchus keta) using a chemiluminescent acridinium ester as a label. The immune complex transfer method dramatically reduced non-specifically bound of acridinium ester labelled antibody without a decrease in the specific binding. Consequently, we could detect lower levels of GH than achieved previously in a two-site CLIA for salmon GH. The detection limit of the assay was 7.8 fg/ml and the standard curve was linear up to 250 fg/ml. Coefficients of variation were 2.2-7.7% within-assay and 5.3-91% between-assay. We have developed a highly sensitive and reproducible GH method and applied it to measurement of GH in alevin chum salmon.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.