Abstract

Immune adherence hemagglutination (IAHA) was compared to complement fixation (CF), using standard procedures, for serological testing of human sera with a number of commercially available antigens. The antigens included herpes simplex, measles, cytomegalo-, and influenza (type B) viruses, as well as Mycoplasma pneumoniae and Chlamydia psittaci (Chlamydia group). The IAHA test was found to be as specific as the CF test, but 4 to 20 times as sensitive with all antigens tested. Antigen titers were also higher with the IAHA method, and the time required to complete the test was only 4 h for the IAHA method, compared with 20 h for the CF method. The increased sensitivity of the IAHA test should permit its use for determination of immunity, as well as for serodiagnosis of recent infections.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.