Abstract
Fish muscle extracts (Scomberomorus— brasiliensis- carite) were analyzed for their hypoxanthine content using a flow injection system incorporating an immobilized xanthine oxidase bioreactor. The xanthine oxidase was immobilized under mild conditions to a 2-fluoro-1-methylpyridinium Fractogel support. The uric acid produced from the oxidation of hypoxanthine by the immobilized xanthine oxidase at pH 7.0 and 35‡C was monitored at 290 ran. Hypoxanthine concentrations as low as 4.4 Μmol/L can be detected. Up to 30 samples per hour can be analyzed at a flow rate of 1 mL/min, using 150 ΜL sample volumes and a bioreactor dimension of 1.0 cm x 2.0 mm id. Recovery yields were between 92 and 99%. Both within day and between day precisions gave CVs < 5.00% (n = 30). Good correlation (r = 0.998) is obtained when 78 fish samples were analyzed for their hypoxanthine content both by this FI method and a reference HPLC method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.