Abstract In low vacuum scanning electron microscopes, the primary beam is partially scattered by the gas present in the specimen chamber. The development of these microscopes, in particular the so-called ‘Environmental’ SEM, was initiated when it was realized that this scattering does not necessarily compromise the imaging capabilities of the instrument. Indeed, some modern commercial instruments are capable of better than 2 nanometer resolution at gas pressures of several torr. The accepted explanation for this is as follows: The mean-free-path of the high energy primary electrons is several millimeters in one torr of water vapour (for example). Because the actual pathlength of electrons travelling through the gas is only a few millimeters, most of them do not scatter at all. Those that do scatter are supposedly distributed over a relatively large area. Thus, the probe features a high-intensity central region surrounded by a slowly decaying low-intensity skirt. High resolution imaging is possible because the signal-to-(skirt)background ratio is high.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.