Abstract

The imaging parameters of non-contrast three-dimensional time-of-flight magnetic resonance angiography (3D TOF-MRA) were optimized to improve the image quality for patients treated using stent-assisted coiling. A simulated blood flow phantom with three types of stents (Enterprise 2, Neuroform Atlas, and LVIS) was imaged by changing echo time (TE), band width (BW), flip angle (FA), and matrix (phase, frequency). The difference between the signal intensity in the simulated vessel and the background was measured at each imaging condition. The ratio of this difference with and without the stent was evaluated as the relative in-stent signal (RIS). In addition, the error ratio of the stent lumen diameter was assessed by comparing the full width at half maximum (FWHM) to that measured by 3D X-ray angiography. The RIS was higher in order of LVIS, Neuroform Atlas, and Enterprise 2 in all conditions. The RIS was higher in imaging conditions with short TE, narrow BW, high FA, and large phase matrix. The highest RIS was seen with a frequency matrix of 320 in the Enterprise 2 and 256 in the others. FWHM error ratio was smaller in the same order as the RIS. FWHM error ratio was smaller in imaging conditions with short TE, large frequency matrix (>384), large phase matrix (>224), and high FA (>20°). Imaging conditions of 3D TOF-MRA that were effective to improve the image quality for stent lumen evaluation were short TE and high spatial resolution.