Abstract

Purpose: Cadmium zinc telluride (CZT) detectors have recently been introduced to the field of clinical nuclear cardiology. However, the feasibility of using them for organs other than the heart remains unclear. The aim of this study was to evaluate the potential of a cardiac CZT camera to acquire thyroid and parathyroid images. We used custom-made phantoms and the currently available standard protocols for CZT, instead of a sodium-iodine scintillation (NaI) camera. Materials and Methods: Thyroid phantoms with or without parathyroid adenomas were made from agar using radiopharmaceuticals (99mTc or 123I) and imaged using CZT and NaI cameras. Using the CZT camera data, we prepared maximum intensity projection (MIP) images and planar equivalent (PE) images. Image counts were compared to those from the NaI camera, and the radioactivity of the phantoms was measured. For parathyroid imaging, three different protocols with the NaI camera were tested using MIP images. Results: For thyroid imaging, MIP could provide images as clear as those obtained from the NaI camera. The radioactivity and image counts correlated better for the PE images than the MIP images, especially for 123I images. We succeeded in obtaining clear parathyroid adenoma images from MIP images using all three protocols. Conclusion: A cardiac CZT camera can effectively perform qualitative and quantitative assessments of the thyroid and parathyroid organs.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call